ClearanceBrief
Recall Z-2108-2025 ยท MLZ Vitrectomy, Instrument Cutter

Alcon Research LLC

Brand Name: Alcon Laboratories, Inc. Product Name: Constellation Vision System Vitrectomy Procedure Pak Model/Catalog Number: Many - See 'Attachment 1 - Affected Product List' Software Version: N/A Product Description: Surgical procedure pack to interface with Alcon Constellation Surgical console. Component: N/A

DateJul 10, 2025 (initiated Apr 28, 2025)
StatusOpen, Classified
ReasonDue to the potential that some units within specific lots were damaged during the manufacturing process. The area of damage is within the same area of the lidding. While the damage in the lidding is visibly noticeable, due to risk of sterility being compromised.
Root causeUnder Investigation by firm
ActionOn 05/01/2025, the firm sent via Overnight FedEx an "URGENT: MEDICAL DEVICE RECALL" Letter to customers informing them that there is a potential for some trays within impacted lots to have damage in the Tyvek cover and compromising the sterility of the products. Customers are instructed: For Standalone Constellation kits- 1. Review inventory to locate, segregate and dispose of affected products 2. Call Alcon Customer Service to arrange for replacement products 3. Forward Post Recall Notice to all department within their organization and to facilities where the product may have been transfer t
Quantity198205
DistributionWorldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV. The countries of Albania, Ar
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K141065CONSTELLATION VISION SYSTEM
Alcon Research, Ltd., Aug 22, 2014