ClearanceBrief
Recall Z-1905-2025 ยท SAB Table Mounted Miniaturized Electromechanical Surgical System

VIRTUAL INCISION CORPORATION

MIRA SURGICAL SYSTEM, Table mounted miniaturized electromechanical surgical system, Catalog Number MIRA-CAP-110, Software Version 4.0.3

DateJun 4, 2025 (initiated Apr 23, 2025)
StatusCompleted
ReasonThe MIRA Surgical System Software Version 4.0.3 loaded on the Surgeon Control Console may have the potential to exhibit interrupted motion on the display screen, caused by a software latency between the movement of the hand controllers and the movement of the Surgical Minibot arms.
Root causeSoftware design
ActionA "MEDICAL DEVICE CORRECTION NOTIFICATION" letter dated 4/22/25 was sent to customers. Actions to be taken by the Customer/User: Do not use the device and transition scheduled procedures to another minimally invasive surgical technique i.e. laparoscopic or other robotic assisted device. Remove the device from service using your internal institutional procedures (e.g. Lock-Out, Tag-Out) for device maintenance until Virtual Incision technicians service the devices on hand. There is no anticipated impact to the conduct of the clinical study under the IDE. Please complete the attached Acknowledg
Quantity3 units
DistributionUS Distribution to States: CA, FL, NE
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.