Recall Z-1892-2025 ยท ITX Transducer, Ultrasonic, Diagnostic

Civco Medical Instruments Co. Inc.

BX2 Needle Guide, Part Number 644-094, tool for performing needle-guided (or catheter) procedures with the use of diagnostic ultrasound transducers

DateMay 30, 2025 (initiated May 1, 2025)
StatusOpen, Classified
ReasonDuring the attachment of the needle guide to the covered probe, the tight tolerance and fit of the needle guide can cause a pinhole breach in the transducer cover when latching the guide in place.
Root causeProcess design
ActionAn URGENT: MEDICAL DEVICE RECALL notification letter dated 5/1/25 was sent to customers. Actions to be taken by the Customer/User: CIVCO requests that you check your inventory for any BX2 Needle Guides and complete the supplied response letter for the appropriate action. Please contact your distribution partner to return any existing inventory or report its destruction by completing the attached response form. Please complete the enclosed response form even if you do not have any BX2 Needle Guides remaining in your inventory to assist in our reconciliation process. Response forms can also b
Quantity506 boxes
DistributionOUS Distribution to countries of: Japan, Taiwan, France, Georgia, Germany and Denmark
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K093713ULTRASOUND GENERAL PURPOSE GUIDANCE SYSTEM
Civco Medical Instruments, Jan 19, 2010