Recall Z-1891-2025 ยท PQP Next Generation Sequencing Oncology Panel, Somatic Or Germline Variant Detection System

Foundation Medicine, Inc.

FoundationOne Companion Diagnostic (F1CDx)

DateMay 30, 2025 (initiated Feb 15, 2023)
StatusOpen, Classified
ReasonAn incorrect negative claim was identified on the claims page; the device variant information was displayed correctly in the tumor profiling section of the FDA-approved test report.
Root causeSoftware design
ActionOn February 15, 2023, the physician was alerted of the amended report within the physician's portal. Physician was informed original report included an incorrect negative CDx claim on the claims page; the CDx variant information was displayed correctly in the tumor profiling section of the FDA-approved test report. It is requested that any impact to patient care be forwarded to Foundation Medicine.
Quantity1 unit
DistributionUS Nationwide distribution in the state of WI.
FDA recordOpen on fda.gov