ClearanceBrief
Recall Z-1865-2025 ยท FRN Pump, Infusion

Zyno Medical LLC

Z-800 Infusion System. Intended to provide intravenous infusion of parenteral fluids, blood, and blood products to a patient.

DateJun 6, 2025 (initiated May 7, 2025)
StatusOpen, Classified
ReasonUnreleased software versions were installed on distributed devices without verification or validation.
Root causeSoftware design (manufacturing process)
ActionConsignees received an "URGENT: MEDICAL DEVICE CORRECTION" notification dated 5/7/25 by email. The notification instructs consignees to check their stock for affected devices and to immediately stop use; recalled devices are to be returned to Zyno Medical via the provided Response Verification Form. The recall notice is to be provided to personnel at consignee locations who need to be informed and communicated to facilities to where devices were transferred to. Any questions about this recall action are to be directed via email to Frederick Lee at feedback@intuvie.com. A representative from Z
Quantity613 units
DistributionUS Nationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K100705ZYNO Z-800 INFUSION SYSTEM MODEL Z-800
Zyno Medical, LLC, Jul 8, 2010
K130690Z-800 INFUSION SYSTEM
Zyno Medical, LLC, Jun 12, 2013