Recall Z-1858-2025 ยท IYE Accelerator, Linear, Medical

C-RAD POSITIONING AB

Brand Name: Catalyst+ Product Name: C4D software used in conjunction with Catalyst (SP001-0025), Catalyst HD (SP001-0026), Catalyst PT (SP001-0027), Catalyst X4 (SP002-0035), Catalyst+ (SP003-0001), Catalyst+ HD (SP003-0002), Catalyst+ PT (SP003-0008), Catalyst+ X4, (SPOO3-0009) Product Description: Patient positioning device, diagnostic imaging/radiotherapy, laser Component: Not a comp

DateMay 28, 2025 (initiated Apr 22, 2025)
StatusOpen, Classified
ReasonSoftware issue with scanning equipment that can results in the filed rotation not being applied correctly.
Root causeSoftware design
ActionOn April 22, 2025, C-RAD Positioning AB issued a Field Safety Notice to affected consignees via E-Mail. C-RAD asked consignees to take the following actions: 1. As an immediate action, the affected users are recommended to: - Do not use Site Groups for treatment plans containing non-coplanar fields. - Ungroup all existing Site Groups. 2. Ensure that all the staff operating the Catalyst family systems and the c4D software application are aware of this information. 3. All customers are requested to reply to C-RAD using the attached FSN Customer Reply Form .
Quantity44 units
DistributionWorldwide - US Nationwide distribution in the states of AK, FL, WA, VA, WA, NC, IL, AR, NC and the countries of Germany, Italy, Uzbekistan, South Korea, Vietnam, Austria, Australia, China, Estonia, Belgium, Uzbekistan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K200435Catalyst+, Catalyst+ HD
C-Rad Positioning AB, May 7, 2020