Recall Z-1988-2025 ยท HCG Device, Neurovascular Embolization

BALT USA, LLC

Optima Coil System REFs/UDI-DI codes: OPTI0156BLK 00810068568875 OPTI0158BLK 00810068568882 OPTI0103BLK 00810068568851 OPTI0258BLK 00810068568929 OPTI0212BLK 00810068568912 OPTI0206BLK 00810068568905 OPTI0310BLK 00810068568943 OPTI0320BLK 00810068568950 OPTI0151HSS10 00818053026614 OPTI0152HSS10 00840303701002 OPTI0153HSS10 00840303701019 OPTI0154CSS10 00818053025761 OPTI0154HSS10 00840303701026 O

DateJun 18, 2025 (initiated May 8, 2025)
StatusOpen, Classified
ReasonDue to radiopaque (RO) marker was not visible during angiography and it was determined that the RO marker was not manufactured to specification.
Root causeVendor change control
ActionOn 05/08/2025, the firm sent via priority mail and an, "URGENT: MEDICAL DEVICE RECALL" Letter informing customers that defect impacting specific lots of the Optima and Prestige Coil Systems, in which the distal 3 cm radiopaque (RO) marker may not be visible under fluoroscopy. Customers are instructed to: As a voluntary alternative to returning the affected products, devices from the affected lots can be inspected by trained hospital personnel for the presence of the RO marker using available fluoroscopy imaging equipment prior to a clinical procedure. It is important to conduct this inspectio
Quantity2889 devices
DistributionWorldwide - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IL, IN, KS, KY, MA, ME, MN, MO, MS, NC, ND, NE, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA and WI. The countries of Argentina, Armenia, Australia, Belgium, Brazil, Bulgaria
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K172390Optima Coil System
Blockade Medical, LLC (D.B.A. Balt Usa), Feb 18, 2018
K200030Optima Coil System
Balt USA, LLC, Feb 1, 2020