Recall Z-1816-2025 ยท KNC Table, Obstetric (And Accessories)

Baxter Healthcare Corporation

Affinity 4 Birthing Bed, Product Codes P3700E100429, P3700E100432, P3700E100433, P3700E100434, P3700E100435, P3700E100436, P3700E100437, P3700E100438, P3700E100439, P3700E100441, P3700E100442, P3700E100443, P3700E100444, P3700E100445, P3700E100446, P3700E100448, P3700E100449, P3700E100450, P3700E100453, P3700E100455, P3700E100456, P3700E100457, P3700E100459, P3700E100463, P3700E100468, P3700E10047

DateMay 23, 2025 (initiated May 6, 2025)
StatusOpen, Classified
ReasonThere is a potential for improper crimps on the signal wires of the actuators which could result in a loss of ability to lower the bed or the foot section of the bed.
Root causeNonconforming Material/Component
ActionAn URGENT MEDICAL DEVICE CORRECTION notification letter dated 5/6/25 was sent to customers. Actions to be Taken by Customers 1. Customers may continue to use the potentially impacted beds until an inspection/correction has been performed by Baxter. 2. If you received this communication directly from Baxter, acknowledge receipt by following the instructions on the enclosed reply instruction sheet, even if you have no remaining inventory. Acknowledging receipt of this notification will prevent you from receiving repeat notices. If you do not complete the acknowledgement, you will receive a phon
Quantity164 units
DistributionWorldwide distribution - US Nationwide and the countries of BAHRAIN, Canada, Chile, FRANCE, KUWAIT, NETHERLANDS, NORWAY, UNITED ARAB EMIRATES.