Recall Z-1913-2025 ยท JLW Radioimmunoassay, Thyroid-Stimulating Hormone

Beckman Coulter, Inc.

DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137, running software version 1.20

DateJun 5, 2025 (initiated Apr 23, 2025)
StatusOpen, Classified
ReasonA new software update (v1.20) is introducing error codes for the Dxl 9000 which prevents the assay from being performed and a delay in results from being generated, and may require field service to resolve the delay.
Root causeSoftware design
ActionAn "URGENT MEDICAL DEVICE RECALL (UMDR)" notification letter dated 4/16/2025 was sent to customers. ACTION: If your laboratory has experienced the red system event and it is not resolved by initializing the analyzer, contact your Beckman Coulter representative and schedule a service visit. RESOLUTION: Beckman Coulter will release a modification for this issue. Your service representative will schedule the modification installation when it is available. Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documenta
Quantity231 units
DistributionWorldwide Distribution: US (Nationwide) and OUS (Global) to countries of: Kazakhstan, Spain, Germany, Italy, Belgium, Croatia, Taiwan, Province of China, Poland, India, Singapore, Switzerland, France, Libya, United Kingdom of Great Britain and Northern Irland, Oman, Algeria, Czechia, Korea, Public o
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K221225Access TSH (3rd IS) Assay, DxI 9000 Access Immunoassay Analyzer
Beckman Coulter, Inc., Nov 10, 2022