Recall Z-1926-2025 ยท JLW Radioimmunoassay, Thyroid-Stimulating Hormone

Beckman Coulter, Inc.

DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137, running software versions 1.20

DateJun 6, 2025 (initiated Apr 23, 2025)
StatusOpen, Classified
ReasonA new motor/encoder and cable was introduced into instruments manufactured starting with serial number 300141 incorrectly. This incorrect installation is introducing error codes for the Dxl 9000 which prevents the assay from being performed and a delay in results from being generated, and may require field service to resolve the delay.
Root causeNonconforming Material/Component
ActionAn "URGENT MEDICAL DEVICE RECALL (UMDR)" notification letter dated 4/16/25 was sent to customers. ACTION: If your laboratory encounters the system event described in this letter: Perform a system initialization to resume normal operation. Repeat testing for any canceled tests. No action is required if your laboratory has not observed the system event listed above. RESOLUTION: Beckman Coulter will release a modification for this issue. Your service representative will schedule the modification installation when it is available. Please share this information with your laboratory staff and
Quantity324 units
DistributionWorldwide distribution: US (Nationwide) and OUS Global (Foreign) to countries of: Italy, Germany, Spain, New Zealand, France, Australia, Slovakia, Switzerland, Ireland, Brazil, Taiwan, Province of China, India, United Kingdom of Great Britain and Northern Ireland, Hungary, Austria, Portugal, Croatia
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K221225Access TSH (3rd IS) Assay, DxI 9000 Access Immunoassay Analyzer
Beckman Coulter, Inc., Nov 10, 2022