ClearanceBrief
Recall Z-1896-2025 ยท GEI Electrosurgical, Cutting & Coagulation & Accessories

Silex Medical, LLC

Signature Laparoscopic Instruments, Sharp Tooth Grasper with Rotating Shaft and Monopolar Electrosurgical Footswitch Connector. Model/REF Number: 5863020.

DateJun 2, 2025 (initiated Jan 9, 2024)
StatusOpen, Classified
ReasonInsulation on the tube did not extend fully to the proximal end of the shaft on the instrument, leaving an exposed (not insulated) area of the shaft closer to the rotator wheel.
Root causeProcess control
ActionSilex formally notified customers on 01/09/2024. Instructions to consignees included to inspect all instruments that they received from the affected lots for the defect. If any instruments had the issue, the customers were instructed to send it back to their facility under an RGA.
Quantity29 units
DistributionUS Nationwide distribution in the states of Florida, North Dakota, Illinois, California, Texas, Virginia, Michigan, Connecticut, Oregon, Iowa, Mississippi, New Jersey, Idaho, West Virginia, Hawaii.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K914910DAVOL LAPAROSCOPIC INSTRUMENTS (GEN SURG REUSABLE)
Davol, Inc., Jan 6, 1992