Recall Z-1754-2025 ยท NVM Catheter, Angioplasty, Peripheral, Transluminal, Dual-Balloon

Numed Inc

BIB Balloon in Balloon Dilatation Catheter. Model Number: 420

DateMay 13, 2025 (initiated Apr 21, 2025)
StatusTerminated
ReasonBalloon extension lumens are not printed correctly. Specifically, the inner balloon inflation hub was labeled as the outer balloon, and the outer balloon inflation hub was labeled as the inner balloon.
Root causeMixed-up of materials/components
ActionNuMED sent a "URGENT: MEDICAL DEVICE RECALL" letter dated 4/21/2025 via email on 4/21/2025 to their single customer. The customer was instructed to immediately return the affected units as soon as possible. If you have any questions or concerns, please contact Regulatory Affairs Manager at (315) 328-4491, or via email at nlaflesh@numedusa.com.
Quantity4 units
DistributionUS distribution to Pennsylvania.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K050857BIB PTA BALLOON CATHETER
NuMED, Inc., Jul 28, 2005