Recall Z-1911-2025 ยท NWB Endoscope, Accessories, Narrow Band Spectrum

Olympus Corporation of the Americas

Olympus EVIS EXERA III Duodenovideoscope. Model Number: TJF-Q190V

DateJun 5, 2025 (initiated May 5, 2025)
StatusOpen, Classified
ReasonRecent post-market surveillance data suggests a possible association of higher microbial contamination levels from TJF duodenoscopes when manual cleaning was delayed beyond one hour and a presoak was performed, compared to those TJF duodenoscopes where manual cleaning began within one hour after patient procedure.
Root causeLabeling design
ActionOlympus notified consignees on 05/05/2025 via "URGENT Medical Device Correction" letter. Consignees were instructed to carefully review the notification, ensure all reprocessing and endoscopy unit personnel are completely knowledgeable and thoroughly trained on the updated reprocessing instructions, replace any copies of the reprocessing manual with the updated version (RC8415_02), complete and return the response form to recall web portal: https://olympusamerica.com/recall, enter the recall number: "0471" and complete the form as instructed, and forward the notification to affected customers
Quantity12,195 units
DistributionWorldwide Distribution: US (Nationwide) and OUS (International) countries to: Algeria, Angola, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Belarus, Belgium, Benin, Bolivia, Brazil, Canada, Cayman Islands, Chile, Colombia, Costa Rica, Croatia, Denmark, Dominican Republic, Egypt, El S
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K220587Evis Exera III Duodenovideoscope Olympus TJF-Q190V
Olympus Medical Systems Corp., May 11, 2022