Recall Z-1767-2025 ยท FRN Pump, Infusion

Baxter Healthcare Corporation

Baxter Novum IQ LVP INFUSION SYSTEM, REF 40700BAXUS

DateMay 20, 2025 (initiated Apr 24, 2025)
StatusOpen, Classified
ReasonThere is a potential for the Novum IQ LVP due to the potential for underinfusion following use of the "standby mode" feature or if the device is powered off with the set loaded.
Root causeDevice Design
ActionBaxter issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 04/24/2025 via USPS First Class Mail. The notice explained the issue, hazard, and requested the following: Actions to be Taken by Customers 1. For flow rates greater than 50 mL/hour, do not exceed a programmed standby time of 2 hours and 30 minutes. Monitor patients frequently to ensure that the appropriate infusion is being delivered. 2. Please remove the set upon powering off the device. 3. Post the enclosed informational poster with Novum IQ LVPs in your facility. 4. Acknowledge receipt of this notificat
Quantity34,524 units
DistributionUS and Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K211122Novum IQ Large Volume Pump
Baxter Healthcare Corporation, Mar 29, 2024