Recall Z-1729-2025 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

Remote Diagnostic Technologies Ltd.

Philips Tempus Pro Patient Monitor, REF: 00-1004-R, 00-1007-R, 00-1024-R, 00-1026-R, 00-1004, 00-1007

DateMay 7, 2025 (initiated Apr 15, 2025)
StatusOpen, Classified
ReasonPatient Monitor software update to address issues:1)May freeze at startup with MPM Application error and Attention Restart Required messages,2)Monitor may freeze and reboot at start of 12-lead ECG or after, 3)May reboot after intermittent connection with Smart Mount, 4) 12-lead ECG measurements with Louvain algorithm and software before vx.36 could produce inaccurate rhythm/morphology statements.
Root causeSoftware design
ActionOn 4/15/2025, correction notices were mailed to customers and distributors informing them of the following: I) The User/Operator Manual provides precautions and warnings on maintaining appropriate power or battery supply, which may reduce the risk of occurrence of an "MPM Application" software error message: -Warning (In Chapter 2, Getting started): Ensure the latches on both sides of the battery are fully engaged prior to using the Tempus - an incorrectly fitted battery could result in the tempus losing power during use. -Note: Before removing the battery, you must switch off the Tempus Pro
Quantity6961
DistributionWorldwide - US Nationwide distribution including in the states of AZ, VA, FL, CA, TX, NM, IA, NJ, TN, OH, KS, IN, NC, IL, WY, WI, NV, NY, NE, PA, AL, , MA, GA, MO, CO, OK, KY, OR, MS, MI, HI, WA, UT, WV, ID, ND, NH, MN, MT, ME, MD, DE, DC and the countries of AU, GB, IT, SA, DE, AE, BO, FR, ES, BE,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K201746Tempus Pro Patient Monitor
Remote Diagnostic Technologies Limited, Sep 18, 2020