Recall Z-1822-2025 · JIS Calibrator, Primary
Siemens Healthcare Diagnostics, Inc.
ADVIA Centaur HCY Calibrator (2 Pack). Siemens Material Number: 10310376.
| Date | May 27, 2025 (initiated Apr 15, 2025) |
| Status | Open, Classified |
| Reason | Certain lots of Atellica IM Homocysteine Calibrator and ADVIA Centaur HCY Calibrator do not meet the assay Instructions for Use (IFU) reference interval due to calibrator instability resulting in increased positive bias. |
| Root cause | Under Investigation by firm |
| Action | An Urgent Medical Device Correction (UMDC) was mailed to all US affected customers by FedEx, and an Urgent Field Safety Notice (UFSN) was issued to all OUS Siemens reporting sites with affected customers on about 04/15/2025. Customers were instructed to review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, discontinue use of and discard the affected devices, and complete and return the Effectiveness Check questionnaire. Customers were also instructed to forward the letter to those who may have received the |
| Quantity | 1,072 packs (2,144 units) - 156 packs US, 916 packs OUS |
| Distribution | Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KY, MA, MD, ME, MI, MO, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, TX, UT, VA, VT, WI, WY and the countries of Austria, Belgium, Croatia, Curacao,¿St¿Eus, Czech¿Republi |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K013406 | ADVIA CENTAUR HOMOCYSTEINE ASSAY Bayer Diagnostics Corp., Dec 18, 2001 |