Recall Z-1821-2025 · JIS Calibrator, Primary

Siemens Healthcare Diagnostics, Inc.

Atellica IM Homocysteine Calibrator (2 Pack). Siemens Material Number: 10995498

DateMay 27, 2025 (initiated Apr 15, 2025)
StatusOpen, Classified
ReasonCertain lots of Atellica IM Homocysteine Calibrator and ADVIA Centaur HCY Calibrator do not meet the assay Instructions for Use (IFU) reference interval due to calibrator instability resulting in increased positive bias.
Root causeUnder Investigation by firm
ActionAn Urgent Medical Device Correction (UMDC) was mailed to all US affected customers by FedEx, and an Urgent Field Safety Notice (UFSN) was issued to all OUS Siemens reporting sites with affected customers on about 04/15/2025. Customers were instructed to review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, discontinue use of and discard the affected devices, and complete and return the Effectiveness Check questionnaire. Customers were also instructed to forward the letter to those who may have received the
Quantity2,212 packs (4,424 units) - 448 packs US, 1,764 packs OUS
DistributionWorldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KY, MA, MD, ME, MI, MO, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, TX, UT, VA, VT, WI, WY and the countries of Austria, Belgium, Croatia, Curacao,¿St¿Eus, Czech¿Republi
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.