ClearanceBrief
Recall Z-2028-2025 ยท JQP Calculator/Data Processing Module, For Clinical Use

Beckman Coulter Inc.

REMISOL Advance UDI-DI code: 13700962601874 All Software Versions: B92487; B92488; C24317; C28652; C37500; D04164; C44703; C57017; C69412; C69413; C73942; C73941; C88470; C88471 The Normand Info REMISOL Advance Data Manager collects and manages data and workflow for 128 connections (Beckman Coulter Instruments, automations, LIS&). It can also be connected to a host system. It provides data manage

DateJun 25, 2025 (initiated Apr 8, 2025)
StatusOpen, Classified
ReasonDue to a software issue, the software may incorrectly process the date of birth (DoB) received from the Laboratory Information System (LIS) and when the DoB is set to 01/01/1900 or left empty, REMISOL Advance may incorrectly identify a patient as a newborn, affecting the execution of validation rules based on incorrect reference ranges. The test results are not changed by this software defect and are accurately displayed by REMISOL Advance; however, the evaluation flagging of the results may be affected. This issue may lead to wrong reference range calculations and can result in erroneous flag
Root causeSoftware design
ActionOn 04/07/2025, the firm sent via email, or postal mail an "Urgent Medical Device Recall" Letter to customers informing them that Beckman Coulter has identified an issue with REMISOL Advance when receiving a sample order from the Laboratory Information System (LIS) if the patient's Date of Birth (DoB) is not entered or is unknown. Customers are instructed to: . Verify that their system is configured with the default validation rules and ranges. REMISOL Advance will utilize these default validation rules in situations where the patient's age or date of birth is unknown (please see the chapter
Quantity1,805 systems
DistributionWorldwide - U.S. Nationwide distribution including in the states of AK, AL,AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. The countries of Australi
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.