Recall Z-1722-2025 ยท CCK Analyzer, Gas, Carbon-Dioxide, Gaseous-Phase

Philips North America Llc

Microstream Advance Neonatal-Infant Intubated CO2 Filter Line marketed as Adt/Pedi Intub CO2 Line STerm , Adt/Pedi Intub CO2 Line STerm Lng , Adt/Pedi Intub CO2 Line High Humidity , Adt/Pedi Intub CO2 Line LTerm Lng , Adt/Pedi Intub CO2 Line LTerm , Neo/Inf Intub CO2 Line LTerm , Neo/Inf Intub CO2 Line High Humidity , Neo/Inf Intub CO2 Line LTerm Lng , VitaLine H Set Adult/Pediatric , VitaLine H S

DateMay 5, 2025 (initiated Apr 7, 2025)
StatusOpen, Classified
ReasonDisconnection of the adapter from the endotracheal tube can potentially be difficult or not possible.
Root causeUnknown/Undetermined by firm
ActionConsignees and Distributors were mailed an URGENT Medical Device Correction notice dated 4/7/25. Consignees were instructed to continue using the devices per the IFU and to ensure during setup that the airway adapter can easily be attached and detached from the breathing circuit/tubing before proceeding. The provided recall notification should be forwarded to those in the organization who use the device or where devices may have been transferred. Consignees are asked to utilize the provided QR code to complete a Response Form digitally; consignees can also email a hard copy Response Form to
Quantity486,631 units
DistributionWorldwide distribution - US Nationwide and the countries of AE, AL, AO, AR, AT, AU, AW, BD, BE, BG, BH, BO, BR, BW, BY, CA, CD, CH, CI, CL, CN, CO, CR, CY, CZ, DE, DK, DO, DZ, EC, EE, EG, ES, ET, FI, FJ, FO, FR, GA, GB, GF, GH, GI, GP, GR, GT, HK, HN, HR, HU, ID, IE, IL, IN, IR, IS, IT, JO, JP, KE,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K113441PHYSIOLOGICAL MONITOR, PATIENT MONITOR
Philips Medizinsysteme Boeblingen GmbH, Cardiac AN, Mar 22, 2012
K181467Microstream Luer Adult-Pediatric Intubated Sampling Line, Microstream Advance Adult-Pediatric Intubated Filter Line
Oridion Medical 1987 , Ltd., Feb 8, 2019
K980324MICROSTREAM FILLERLINE OR/EMS
Oridion Medical , Ltd., Apr 16, 1998
K980327MICROSTREAM FILTERLINE ICU
Oridion Medical , Ltd., Apr 16, 1998
K993383HP VIRIDIA MULTI-MEASUREMENT SERVER/COMPACT PORTABLE PATIENT MONITOR MODELS M3000A, M3046A, M3015A, REL.B
Hewlett-Packard GmbH, Feb 3, 2000