Recall Z-1987-2025 ยท NHA Abutment, Implant, Dental, Endosseous

Preat Corp

Biomet 3i Certain-compatible 4.1 mm Milled Ti Abutment REF 9001767-F Preat milled Abutments are intended to be used in conjunction with endosseous dental implants in the maxillary or mandibular arch to provide support for single-unit or multi-unit prosthetic restorations.

DateJun 18, 2025 (initiated Apr 2, 2025)
StatusOpen, Classified
ReasonDue to manufacturing issues, abutment products were manufactured with screw seat located too high, which may cause the mating screw to have a shorter engagement length into the associated implant.
Root causeProcess control
ActionOn 04/02/2025 an initial customer notification was distributed by the firm. On 05/05/2025, the firm sent via certified mail an updated "URGENT: MEDICAL DEVICE RECALL UPDATE" Letter to customers informing them that due to manufacturing issues with a dimensional condition that may lead to no engagement between the prosthetic screw and the implant. Customer are instructed to discard or return the affected devices to PREAT Corporation. For questions, contact Quality Assurance at 800-232-7732 or email mgonzales@younginnovations.com
Quantity145 abutments
DistributionU.S. Nationwide distribution in the states of AZ, CA, GA, MO, NJ, NY, OH, PA, RI, TN,TX, and WA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K183518Preat Abutments
Preat Corporation, Mar 18, 2019