Recall Z-1660-2025 ยท JAK System, X-Ray, Tomography, Computed

Philips North America Llc

Philips Brilliance CT Big Bore, Software Version 4.8.0.10430. Computed Tomography X-Ray System.

DateApr 23, 2025 (initiated Apr 1, 2025)
StatusOpen, Classified
ReasonUnintended motion issues related to Interventional Control that may lead to collision of the Gantry/table with the operator or patient.
Root causeSoftware design
ActionAn "URGENT Medical Device Correction" Notice dated 4/1/25 was mailed to consignees. The notice instructs consignees on how to identify if their unit is affected by this recall. Consignees are instructed to continue use of affected devices by following the recommended actions outlined in the notice and Appendix A. Consignees are asked to return the completed URGENT Medical Device Correction Response Form within 30 days of notification of this recall event via email: Philips.Recall@philips.com. The recall notification is to be circulated among users and retained near affected devices until the u
Quantity630 units
DistributionWorldwide distribution: US (Nationwide); and OUS (Foreign) to countries of: Algeria, Argentina, Australia, Austria, Belgium, Bolivia, Brazil, Canada, Chile, China, Czech Republic, Estonia, France, Germany, Honduras, Hong Kong, India, Indonesia, Iran, Iraq, Israel, Italy, Lebanon, Libya, Luxembourg,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K171850Philips CT Big Bore
Philips Medical Systems (Cleveland), Inc., Nov 9, 2017