Recall Z-1738-2025 · PHX Shoulder Prosthesis, Reverse Configuration

Tornier S.A.S.

Blueprint Software, BPUE001. Intended for use by surgeons as a pre surgical planner for shoulder replacement.

DateMay 8, 2025 (initiated Apr 10, 2025)
StatusOpen, Classified
ReasonA software bug in Blueprint version 4.2.1 prevents users from accessing cases initiated in previous versions of the software.
Root causeSoftware Design Change
ActionOn April 10, 2025, the firm began notifying customers via Urgent Medical Device Recall letters. Customers were informed that an Online Management System (OMS) software patch was released to address the bug on March 25, 2025. The issue can no longer occur with the introduction of this patch. No actions are required on the customer's part to implement the patch. Version 4.2.2 is currently in development and is scheduled for release by end of Quarter 2, 2025. This patch will implement a permanent fix for the issue. Customers will be prompted to complete this update within the Blueprint Software
Quantity438 users
DistributionBlueprint Version 4.2.1 is only currently available in the US. Domestic distribution nationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K211359Tornier Perform™ Patient-Matched Primary Reversed Glenoid and BLUEPRINT™ Patient Specific Instrumentation
Tornier, Inc., Nov 12, 2021