Recall Z-1723-2025 · EOQ Bronchoscope (Flexible Or Rigid)

Ambu Inc.

Ambu aScope" 5 Broncho HD 5.6/2.8 Sampler Set. Model Number: 622002000US

DateMay 6, 2025 (initiated Apr 7, 2025)
StatusOpen, Classified
ReasonIncorrect labeling in which the front red pouch label did not match the actual size of the medical device. The label on the front of the scope incorrectly states 5.0/2.2 while the back label, shipping box and product inside correctly states 5.6/2.8.
Root causeProcess control
ActionAmbu notified sales representative first on 04/07/2025 with an "Urgent Field Safety Notice" letter and they are instructed to visit the affected accounts to notify them of the recall. The letter instructs customers to identify and quarantine any affected units on hand, to arrange for the return of affected units or destroy them on site, and to complete and return the response form.
Quantity160 units
DistributionUS Nationwide distribution in the states of CA, CO, FL, GA, KY, MD, MT, NJ, NV, PA, TN, TX, VA.
FDA recordOpen on fda.gov