Recall Z-1703-2025 ยท IYE Accelerator, Linear, Medical

IBA Dosimetry GmbH

myQA iON; Article Number: MQ10-000;

DateApr 30, 2025 (initiated Apr 1, 2025)
StatusOpen, Classified
ReasonWrong analysis results can occur if the user creates two structures with the same name, except for one differing by a special character.
Root causeSoftware design
ActionOn April 1, 2025, FIELD SAFETY NOTICE letters were sent to customers. Actions to be taken to prevent the error from occurring: When naming structures, please refrain from using any of the following special characters: % \ / : * ? < > | + Additionally, during the export of files into myQA iON, some TPSs cut off the structure names due to their length. Users are not notified when names are automatically shortened by the system. Please check that the structure name was correctly imported. Distribution of this Field Safety Notice: Please distribute this notice to all those who need to be aware w
Quantity84 units (9US, 75 OUS)
DistributionWorldwide - US Nationwide distribution in the states of AL, AR, FL, GA, KS, MD, OK and the countries of Algeria, Austria, Belgium, Czech Republic, Denmark, France, Germany, Ireland, Italy, Morocco, Norway, Poland, Slovakia, Spain, Israel, Kazakhstan, Kosovo, Libya, South Africa, Turkey, United Kingd
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K201798myQA iON