ClearanceBrief
Recall Z-1700-2025 ยท JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer

Waldemar Link GmbH & Co. KG (Mfg Site)

Posterior Femoral Augment; Item Numbers: (1) 880-310/11 (Sz. 0 - cemented sm), (2) 880-310/21 (Sz. 0 - cemented lg), (3) 880-311/11 (Sz. 1-2 - cemented sm), (4) 880-311/21 (Sz. 1-2 - cemented lg), (5) 880-313/11 (Sz. 3-4 - cemented sm), (6) 880-313/12 (Sz. 3-4 - cemented md), (7) 880-313/21 (Sz. 3-4 - cemented lg), (8) 880-313/22 (Sz. 3-4 - cemented x-lg), (9) 880-315/11 (Sz. 5-6 - cemented sm), (

DateApr 30, 2025 (initiated Apr 3, 2025)
StatusOpen, Classified
ReasonThe device was delivered with a preassembled Femoral Augment screw that was missing its thread.
Root causeNonconforming Material/Component
ActionOn April 3, 2025 URGENT: MEDICAL DEVICE CORRECTION letters were emailed to customers. Actions to be taken by the Customer/User: Please take note of the following instructions for replacing any screws found missing the thread. No product return is required. Instructions for disassembling and replacing a defective screw: Disassembly 1. Use the 2.5 mm Torque Wrench article REF (15-2545) to unscrew the screw from the Augment while pushing the tip of the screw towards the Augment. 2. Exchange the screw. Open an alternative Augment and use its screw. Alternative sizes of Augments are always availab
Quantity908 units
DistributionUS Nationwide distribution in the states of AL, AZ, CA, CO, FL, GA, IN, IL, KS, LA, MD, MO, NV, OH, TX, VA, WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K202924LinkSymphoKnee System
Waldemar Link GmbH & Co. KG, May 13, 2021