Recall Z-1721-2025 ยท MEC Disinfector, Medical Devices

Getinge Disinfection Ab

Washer Disinfector Aquadis 56; Model Number: 56M

DateMay 5, 2025 (initiated Apr 2, 2025)
StatusOpen, Classified
ReasonThe frequency inverter is not programmed by supplier according to specification which causes the circulation pump to start earlier than expected, resulting in additional wear and tear, earlier than expected, on the axial seal in the pump housing.
Root causeUnder Investigation by firm
ActionOn April 2, 2025 URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. Customer Actions 1. Check the serial number of yourWasher Disinfector Aquadis 56 to determine if it is affected. The serial number can be found on the marking label, located on the inside of the front Panel door. 2. Please ensure this message is forwarded to any individuals that need notification within your organization or any organization where the affected devices have been transferred. 3. Complete and return the enclosed response form Getinge by e-mailing a copy to recallresponses.qrc@getinge.com. Note: If y
Quantity424 units (79 US, 345 OUS)
DistributionWorldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Bangladesh, Brazil, Canada, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, I