Recall Z-1739-2025 ยท PHL Powered Exoskeleton

WANDERCRAFT SAS

Atalante X, Powered Exoskeleton. Atalante X is intended to perform ambulatory functions and mobility exercises, hands-free, in rehabilitation institutions under the supervision of a trained operator.

DateMay 8, 2025 (initiated Apr 3, 2025)
StatusOpen, Classified
ReasonPotential for exoskeleton to lose lateral balance and cause patient to fall.
Root causeUse error
ActionAn URGENT SAFETY NOTICE dated 4/3/25 was mailed to consignees. The letter notifies consignees that WANDERCRAFT will be reinforcing their training program for the Atalante X and providing additional instructions by 5/1/25. Until the additional instructions can be provided, users are being instructed to only have certified Atalante X operators maintain the stability of the exoskeleton at all times by keeping at least one hand on the handles throughout the session. Consignees are to share the recall notification with users at their facility and retained for awareness.
Quantity3
DistributionDomestic: PA & NY.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K221859Atalante
Wandercraft SAS, Dec 29, 2022