Recall Z-1807-2025 ยท LLZ System, Image Processing, Radiological

Echopixel, Inc.

TRUE 3D VIEWER SYSTEM Model Number EP-00001-XX Software versions 1.0, 1.25, and 1.35 The t3D-Viewer is designed for use by health care professionals HCPs and is intended to assist the clinician who is responsible for making all patient management decisions. The True 3D Viewer (t3D-Viewer) system is comprised of a commercial off the shelf hardware platform and a proprietary software application

DateMay 21, 2025 (initiated Apr 11, 2016)
StatusOpen, Classified
ReasonImaging system data may be displayed with an incorrect orientation (sidedness) where the patient side is not correct (left, right, superior, inferior, anterior or posterior). This may lead to an inaccurate measurements.
Root causeSoftware design
ActionOn 03/25/2026, the firm sent via email and telephone an Urgent Field Safety Notice, informing customers of a potential misdiagnosis or incorrect surgical treatment planning approach can result when utilizing Enhanced Multi-frame DICOM image data sets. Customers were instructed to: verify if Enhanced Multi-Frame CT DICOM files have been used with the software. If users have utilized this file type, we request that users verify retrospectively any potential risk to health of patients and further determine if this patient needs to be reevaluated and report it to EchoPixel, Inc. We request that t
Quantity9 systems
DistributionWorldwide - U.S. Nationwide distribution in the states of CA, OH, NJ, PA, MA, FL and the country of Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K142107ECHO TRUE 3D VIEWER
Echopixel, Inc., Jan 21, 2015