Recall Z-1668-2025 ยท FRN Pump, Infusion

Smiths Medical ASD, Inc.

CADD-Solis VIP Ambulatory Infusion Pump; Model/Number: 21-2120-0100-02 21-2120-0100-03 21-2120-0100-06 21-2120-0100-232 21-2120-0100-233 21-2120-0100-50 21-2120-0100-51 21-2120-0100-95 21-2120-0102-02 21-2120-0102-03 21-2120-0102-06 21-2120-0102-07 21-2120-0102-08 21-2120-0102-12 21-2120-0102-13 21-2120-0102-14 21-2120-0102-15 21-2120-0102-17 21-2120-0102-237 21-2120-0102-238 21-2120-0102-247 21

DateMay 7, 2025 (initiated Apr 10, 2025)
StatusOpen, Classified
ReasonAffected pumps may trigger an erroneous (false) Upstream Occlusion Alarm under certain conditions, which will interrupt an active infusion. Interruption or delay of therapy can lead to serious patient injury or death.
Root causeLabeling design
ActionOn April 10, 2025, the firm notified affected consignees via letters titled "Urgent Medical Device Correction." Customers were informed of conditions under which the false Upstream Occlusion Alarm may occur. The USO alarm can be cleared during these conditions by removing the administration from the pump, re-attaching the administration set, and restarting delivery. Customers should verify there is no upstream occlusion present. The firm will be making an update to the Infusion Pump Operator's Manual to include these conditions.
Quantity185,975 US; 127,701 OUS Total
DistributionWorldwide - US nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K111275CADD SOLIS VIP AMBULATORY INFUSION PUMP AND CADD SOLIS MEDICATION SAFETY SOFTWARE
Smiths Medical Asd, Inc., Feb 1, 2013