Recall Z-1731-2025 ยท KGZ Accessories, Catheter
Galt Medical Corporation
GALT Centeze Catheter REF: DRC-002-05 DRC-002-06 The Centeze is intended for use in percutaneous fluid aspirations and small volume drainage procedures. For both adult and adolescent groups.
| Date | May 7, 2025 (initiated Mar 27, 2025) |
| Status | Open, Classified |
| Reason | Due to a potential open seal in the sterile barrier packaging . |
| Root cause | Process control |
| Action | On 03/27/2025, the firm emailed customers an "URGENT: MEDICAL DEVICE RECALL" Letter informing them that affected products may have an open seal in the sterile barrier packaging that will render the product non-sterile and will increase the risk of patient harm due to microbial and environmental contamination. Customers are instructed to: 1.DO NOT USE OR DISTRIBUTE THE DEVICEs LISTED IN THIS RECALL. Identify and segregate the affected lot(s) that are in their possession as well as those in the possession of their end-users. 2. Complete the enclosed Field Correction Reply Forms and email or fax |
| Quantity | 2,881 catheters |
| Distribution | Worldwide - U.S. Nationwide distribution in the states of AL, FL, GA, IL, MA, OH, PA, and UT. The countries of Canada, Ireland, Israel, Korea, Qatar, and United Kingdom. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.