ClearanceBrief
Recall Z-1731-2025 ยท KGZ Accessories, Catheter

Galt Medical Corporation

GALT Centeze Catheter REF: DRC-002-05 DRC-002-06 The Centeze is intended for use in percutaneous fluid aspirations and small volume drainage procedures. For both adult and adolescent groups.

DateMay 7, 2025 (initiated Mar 27, 2025)
StatusOpen, Classified
ReasonDue to a potential open seal in the sterile barrier packaging .
Root causeProcess control
ActionOn 03/27/2025, the firm emailed customers an "URGENT: MEDICAL DEVICE RECALL" Letter informing them that affected products may have an open seal in the sterile barrier packaging that will render the product non-sterile and will increase the risk of patient harm due to microbial and environmental contamination. Customers are instructed to: 1.DO NOT USE OR DISTRIBUTE THE DEVICEs LISTED IN THIS RECALL. Identify and segregate the affected lot(s) that are in their possession as well as those in the possession of their end-users. 2. Complete the enclosed Field Correction Reply Forms and email or fax
Quantity2,881 catheters
DistributionWorldwide - U.S. Nationwide distribution in the states of AL, FL, GA, IL, MA, OH, PA, and UT. The countries of Canada, Ireland, Israel, Korea, Qatar, and United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.