ClearanceBrief
Recall Z-1674-2025 ยท LGZ Warmer, Thermal, Infusion Fluid

Belmont Instrument LLC

Hyperthermia Pump Procedure Kit. Model Number: 902-00045. Sterile, single-use, disposable set used with the Hyperthermia Pump. The Hyperthermia Procedure kit (P/N 902-00045) consists of 4.4 Liter Reservoir (P/N 902-00034P), Heat Exchanger set (P/N 902-00006P) and HP Procedure Pack (P/N 902-00048P). The three components are packaged separately in individual pouches and boxes and placed in the s

DateApr 25, 2025 (initiated Mar 31, 2025)
StatusOpen, Classified
ReasonPotential crack in the female connector located on the Heat Exchanger in the set, which will lead to a fluid leak through the crack during priming of the rapid infuser.
Root causeProcess design
ActionBelmont Medical notified consignees on about 03/31/2025 via "URGENT MEDICAL DEVICE CORECTION" FedEx or emailed letter. Consignees were instructed to locate affected product and to return the customer Response Form indicating if any affected units were on hand. Consignees were also instructed to follow instructions provided in the letter to inspect affected lots and if any units are identified as having cracks or leaks during priming, they should be discarded. Lot number of the set should be noted and reported to Belmont Medical Technologies. Belmont has indicated that they will replace any
Quantity158 sets
DistributionWorldwide - US Nationwide distribution in the states of AZ, CA, CO, CT, DC, IL, IN, KS, KY, MA, ME, MN, NC, ND, NJ, NM, OH, OR, PA, SC, TN, TX, VA, VT, WA, WI, WV and the countries of Canada, Hungary, Korea (the Republic of), Latvia, Malaysia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K152208The Belmont Hyperthermia Pump
Belmont Instrument Corporation, Sep 2, 2015