Recall Z-1680-2025 ยท MFK Lens, Multifocal Intraocular

Bausch & Lomb Surgical, Inc.

Intraocular Lens. enVista Envy IOL, All models starting with EN; enVista Envy Toric IOL, All models starting with ETN.

DateMay 2, 2025 (initiated Mar 27, 2025)
StatusOpen, Classified
ReasonIn response to an increased number of reports of toxic anterior segment syndrome (TASS) in conjunction with implantation of enVista Aspire and enVista Envy IOLs, and certain enVista monofocal lenses in the U.S.
Root causeProcess control
ActionOn 3/27/2025, a news release was published informing customers that the firm will send out detailed instructions for what to do with any stock on hand. On 4/4/2025, the news release was updated and customers were also informed in a recall notice of the following: For Patients: Cataract surgery patients experiencing eye discomfort should contact their eyecare professional immediately. For Eyecare Professionals: Continue to closely monitor cataract surgery patients for at least 48 hours after surgery. Please instruct patients with new or worsening symptoms to contact your office immediately.
Quantity48,237
DistributionWorldwide distribution. US Nationwide including PR and GU, and the countries of CAN, AUS, AUT, BEL, BGR, IC, HRV, CZE, FRA, DEU, GRC, HUN, ITA, NLD, POL, PRT, ROU, ESP, GBR, NZL, HKG.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.