Recall Z-1683-2025 ยท IEO Processor, Tissue, Automated

Leica Biosystems Melbourne Pty Ltd

HistoCore PELORIS 3, Model/Catalog Number: 45.0005/45.7512.501 A11, Software Version: 3.4.0. The HistoCore PELORIS 3 Rapid Tissue Processor is a dual retort rapid tissue processor used to prepare tissue samples.

DateApr 28, 2025 (initiated Apr 2, 2025)
StatusOpen, Classified
ReasonThere is a leakage issue associated with the tubing in the manifold of the instrument.
Root causeRelease of Material/Component prior to receiving test results
ActionAn URGENT MEDICAL DEVICE RECALL notification letter dated 3/31/25 was sent to customers. ACTIONS REQUIRED 1. If leakage from the instrument is observed in the drip tray or surrounding the instrument, please follow your laboratory's leakage containment/safety protocol. If a protocol is running, wait for the run to finish, switch off the instrument and contact your local Leica Biosystems representative immediately. 2. If no leakage has been observed from the instrument, continue using it as normal. Conduct routine monitoring for leakage until the on-site inspection is completed. 3. Your local L
Quantity77 units
DistributionUS States: AZ, CA, FL, MN, NJ, NY, NC, OH, PA, TN, TX, UT, VA, and WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.