Recall Z-1664-2025 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Beckman Coulter, Inc.

Beckman Coulter DxI 9000 Access Immunoassay Analyzer

DateApr 24, 2025 (initiated Mar 27, 2025)
StatusOpen, Classified
ReasonWhen DxI 9000 Access Immunoassay Analyzer is connected to a host system (laboratory information system (LIS) or middleware) and has accumulated canceled QC test results that were not sent by LIS, the analyzer may lose communication with the host system. The lost connection interrupts sample processing, which delays reporting patient test results and may subsequently delay patient treatment.
Root causeSoftware design
ActionBeckman Coulter issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 03/27/2025 via USPS first class mail. The notice explained the issue with the device, impact to the patient, and provided a workaround until a software upgrade is available. If the consignee further distributed the affected device, they were directed to provide a copy of the notice to the recipient. For questions, contact Customer Support Center, from the website: http://www.beckmancoulter.com
Quantity370 units
DistributionWorldwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K221225Access TSH (3rd IS) Assay, DxI 9000 Access Immunoassay Analyzer
Beckman Coulter, Inc., Nov 10, 2022