ClearanceBrief
Recall Z-1649-2025 ยท LZO Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented

Zimmer, Inc.

Zimmer Biomet Ceramic Head, 22.2 mm Diameter, Type 1, +3 Neck, Model Number 802802203; Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented or Non- Porous, Uncemented

DateApr 21, 2025 (initiated Mar 24, 2025)
StatusOpen, Classified
ReasonThe associated product labeling does not adequately provide compatibility information for use of the 22.2 mm Diameter Ceramic Femoral Heads with cobalt-chrome femoral stems or stainless-steel femoral stems.
Root causeUnder Investigation by firm
ActionZimmer Biomet issued an URGENT MEDICAL DEVICE RECALL CORRECTION - PHASE 1 notice to its consignees on 03/24/2025 via email and courier. The notice explains the issue, potential risk, and requested the following: Distributors: Your Responsibilities 1. Review this notification and ensure that affected team members are aware of the contents. 2. Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. 3. Return the Additional Accounts form to CorporateQuality.PostMarket@zimmerbiomet.com. a. Review the list of hospitals and/or surgeons i
Quantity160 units
DistributionWorldwide distribution - US Nationwide and the countries of Panama, Hong Kong, China.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K181171Zimmer Biomet Ceramic Heads
Zimmer, Inc., Apr 2, 2019