Recall Z-1555-2025 ยท IYN System, Imaging, Pulsed Doppler, Ultrasonic

GE Medical Systems, LLC

Venue Go Models: R2 (Model Number H45181VG), R3 (Model Number H45193VG), R4 (Model Number H45114VG), R5 (Model Numbers H45135VG, H45135VGW), R3 Surgery for China (Model Number H45113VGSU), R3 PeriOP for China (Model Number H45113VGPO), R3 Expert for China (Model Number H45113VGEX ), R3 Pro for China (Model Number H45113VGPR), R3 Performance for China, R3 Intervention for China, R3 Surgery for Chin

DateApr 11, 2025 (initiated Mar 6, 2025)
StatusOpen, Classified
ReasonThe battery for certain ultrasound systems can potentially develop an internal failure which could result in smoke or fire.
Root causeUnder Investigation by firm
ActionAn URGENT MEDICAL DEVICE CORRECTION notification letter dated 3/6/25 was sent to customers. Actions to be taken by Customer/User: You can continue to use your Venue Go and Venue Fit system after following the actions below: 1) Turn the system Off and unplug the system from the wall socket. 2) Remove the system batteries by following instructions provided in the User Manual Chapter 3. These instructions are also provided in this letter in Appendix A for Venue Go and in Appendix B for Venue Fit. 3) Check the date of manufacture of the system batteries by looking at the label on the bottom of
Quantity12,808 total devices
DistributionWorldwide - US Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K183362Venue GO
GE Medical Systems Ultrasound and Primary Care Diagnostics, Jan 28, 2019
K202233Venue Go
GE Medical Systems Ultrasound and Primary Care Diagnostics, Nov 18, 2020
K220800Venue Go
GE Medical Systems Ultrasound and Primary Care Diagnostics, Jun 21, 2022
K240053Venue Go
GE Medical Systems Ultrasound and Primary Care Diagnostics, Jun 10, 2024