Recall Z-1573-2025 ยท OCC Respiratory Virus Panel Nucleic Acid Assay System

Abbott Diagnostics Scarborough, Inc.

Brand Name: ID NOW RSV 24T Product Name: ID NOW RSV 24T Model/Catalog Number: 435-000 Software Version: N/A Product Description: Respiratory virus panel nucleic acid assay system Component: No

DateApr 15, 2025 (initiated Mar 4, 2025)
StatusOpen, Classified
ReasonThe impacted lots have a higher occurrence of invalid rates when compared to the product Instructions for Use.
Root causeProcess change control
ActionOn March 4, 2025 FIELD CORRECTION NOTICE letters were sent to end user customers. Please complete the following actions, as applicable. 1. If you have impacted inventory in stock, then: " Discontinue use of and destroy any remaining inventory of the impacted lots according to your procedures. " Complete and return the Customer Reply Form (Form must be completed, signed and returned to receive a replacement lot). " Please retain this letter for your records. 2. If you have forwarded the product listed above to others in your network, then: - Inform them of this Field Correction Notice, provide
Quantity178056 eaches (147472 US, 31584 OUS)
DistributionWorldwide distribution - US Nationwide and the countries of Australia; Austria; Belgium; Bulgaria; Canada; Costa Rica; France; French Polynesia; Germany; Italy; Luxembourg; Malaysia; Netherlands; Paraguay; Poland; Qatar; Saudi Arabia; Slovenia; South Korea; Spain; Switzerland; Taiwan; United Kingdom
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K161375Alere i RSV, Alere i Instrument, Alere i RSV Control Swab Kit
Alere Scarborough, Inc., Aug 18, 2016