ClearanceBrief
Recall Z-1574-2025 ยท QWR Simple Point-Of-Care Device To Detect Sar-Cov-2 Nucleic Acid Targets From Clinical Specimens In Near-Patient Settings

Abbott Diagnostics Scarborough, Inc.

Brand Name: ID NOW COVID-19 2.0 24T Product Name: ID NOW COVID-19 2.0 24T Model/Catalog Number: 192-000 Software Version: N/A Product Description: Simple Point-Of-Care Device To Detect Sar-Cov-2 Nucleic Acid Targets Component: No

DateApr 15, 2025 (initiated Mar 4, 2025)
StatusOpen, Classified
Reasonthe impacted lots have a higher occurrence of invalid rates when compared to the product Instructions for Use.
Root causeProcess change control
ActionOn March 4, 2025 PRODUCT REMOVAL NOTICE letters were sent to distributors. Necessary Actions to be Taken by Distributor: 1. If you have impacted inventory held in stock, then: "Immediately quarantine impacted product per your local procedure. "Complete and return the Distributor Acknowledgment Form (Form must be completed, signed and returned to receive replacement product). "Contact Customer Service to arrange return of your product by Abbott. 2. If you have forwarded the product listed above to others within your distribution network, then: "Inform them of this Product Removal Notice and pr
Quantity2,164,872 eaches
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K221925ID NOW COVID-19 2.0
Abbott Diagnostics Scarborough, Inc., Aug 10, 2023