ClearanceBrief
Recall Z-1502-2025 ยท GJS Test, Time, Prothrombin

Fisher Diagnostics

Pacific Hemostasis Thromboplastin-D, 4 ml. Used in performing one-stage prothrombin time test and factor assays.

DateMar 31, 2025 (initiated Mar 6, 2025)
StatusOpen, Classified
ReasonThe incorrect value of the International Sensitivity Index (ISI) is printed on the outer box label.
Root causeEmployee error
ActionA "Recall/Field Safety Corrective Action" letter dated 3/5/25 was mailed to consignees instructing them to stop using and discard any affected product in stock, contact their prescribing physicians/Medical Directors for evaluation for further action, and retain a copy of the notification for laboratory records. Consignees are to return the provided Response Form. Consignees with any questions are to contact their Fisher Diagnostic representative or call 800-528-0494. Distributors are to email techsupport.diagnostics.mtn@thermofisher.com with "CDR-FSN-2025-001" in the subject line to coordina
Quantity4,387 units
DistributionUS Distribution to: PA and OUS Foreign country of: United Arab Emirates.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K994100PACIFIC HEMOSTASIS THROMBOPLASTIN D
Pacific Hemostasis, Feb 18, 2000