Recall Z-1559-2025 · MWI Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms)

Baxter Healthcare Corporation

Welch Allyn CONNEX Accessory Power Management Stand:

DateApr 14, 2025 (initiated Mar 24, 2025)
StatusOpen, Classified
ReasonThere were customer reports of devices which experienced battery-related fires. The investigation found that the devices in question were utilizing third-party, after-market batteries that are not approved by Baxter.
Root causeProcess design
ActionAn URGENT MEDICAL DEVICE CORRECTION notification letter dated 3/24/25 was sent to customers. Actions to be Taken by Customers 1. DO NOT USE UNAPPROVED BATTERIES. Baxter asks customers to only use the approved Baxter / Welch Allyn BATT99 battery, consistent with product labeling due to their safety and performance testing. 2. INSPECT ALL DEVICES. If you have a device identified in the affected product table above, check the device for the correct battery type. See Figure 1 below for examples of labels from approved batteries labelled with the BATT99 product code, Welch Allyn , and the Skaneat
Quantity114,202 units
DistributionWorldwide - US Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K241411Welch Allyn Connex® Spot Monitor; 901058 Vital Signs Monitor Core (CSM)
Welch Allyn, Inc., Dec 20, 2024