ClearanceBrief
Recall Z-2477-2025 ยท BRY Cabinet, Table And Tray, Anesthesia

CareFusion 303, Inc.

BD Pyxis Product Name / UDI-DI code / Catalog No. BD Pyxis" MedStation" 4000 Main / 10885403512629 / 303 BD Pyxis" MedStation" ES Tower / 10885403512674 / 352 BD Pyxis" Anesthesia Station ES / 10885403477836 / 327 BD Pyxis" MedStation" ES (Main) / 10885403512667 / 323 BD Pyxis" CII Safe ES / 10885403512605 / 1116-00 BD Pyxis" Enterprise Server / 10885403518348 and 10885403520341 / 1115-00 BD

DateAug 28, 2025 (initiated May 8, 2025)
StatusOpen, Classified
ReasonAntivirus software was not consistently installed on impacted devices during the implementation process.
Root causeUnder Investigation by firm
ActionOn 05/08/2025, the firm sent via mail and email an "URGENT MEDICAL DEVICE PRODUCT CORRECTION" Letter informing customers of two issues. For this recall the issue pertains to technicians failing to install (ESET) antivirus software on affected devices. This can potentially result in a malicious actor that could exploit a cybersecurity weakness on a device that is configured without antivirus protection within a hospital network. Customer are instructed and/or informed: For Clinical Users: -to utilize user facility policies and procedures for cybersecurity issues or disaster recovery. -to remo
Quantity8353 devices (6738 U.S. and 1615 O.U.S.)
DistributionWorldwide - U.S Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MP, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.