Recall Z-1543-2025 ยท LGW Stimulator, Spinal-Cord, Totally Implanted For Pain Relief

Medtronic Neuromodulation

Stimulation RC Clinician Programmer Application, Model A71400, when used with the Inceptiv (Model 977119) and Intellis Pro (Model 977118) neurostimulators

DateApr 8, 2025 (initiated Mar 4, 2025)
StatusOpen, Classified
ReasonThere is a software issue that can permanently disable communication with an implantable neurostimulator (INS) when a specific intraoperative programming sequence is used.
Root causeUnder Investigation by firm
ActionAn "URGENT: MEDICAL DEVICE CORRECTION" notification letter dated March 2025 was sent to customers. The letter described the product, problem and actions to be taken. Required Actions: When utilizing any of the lead options mentioned above, avoid using the Switch Device feature to transfer data from the Model A71300 Stimulation Trialing CP App to the Model A71400 Stimulation RC CP App. As an alternative, manually enter the patient and lead information using the Model A71400 Stimulation RC CP App, ensuring not to utilize the Switch Device feature. Inform relevant medical staff about the potenti
Quantity1493 devices
DistributionWorldwide Distribution: US (Nationwide) and OUS (Global) to countries of: Austria, Belgium, Bulgaria, Canary Islands, Cyprus, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Japan, Luxembourg, Malta, Netherlands, Northern Ireland, Norway, Poland,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.