Recall Z-1553-2025 ยท KPR System, X-Ray, Stationary

Philips North America Llc

Philips DXR System, DuraDiagnost 4.1. Stationary X-Ray System.

DateApr 11, 2025 (initiated Mar 10, 2025)
StatusOpen, Classified
ReasonPotential for collimator to fall as a result of incorrect installation.
Root causeUnder Investigation by firm
ActionConsignees received an URGENT MEDICAL DEVICE CORRECTION NOTICE dated 3/7/25. The notice instructs customers of safety precautions to take during scan preparation to ensure devices are fit for use. If consignees notice that the laser from the collimator or crosshairs of the collimator's light field is not able to align with the silkscreen on the front cover of the detector, consignees are to immediately stop use and call their Philips representative. The provided notice is to be posted near affected systems and circulated to all device users. Philips will reach out to consignees to schedule an
Quantity83 units
DistributionWorldwide - US Nationwide distribution in the states of AK, FL, KS, MS, NH, NY, OR, PA, TX, WI and the countries of Argentina, Australia, Canada, Chile, China, Germany, Hungary, India, Indonesia, Italy, Kenya, Mexico, Netherlands, New Zealand, Oman, Pakistan, Peru, Poland, Romania, Russian Federatio
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K141381DURADIAGNOST
Philips Healthcare (Suzhou) Co., Ltd., Jun 12, 2014
K201640DuraDiagnost
Philips Healthcare (Suzhou) Co., Ltd., Jul 9, 2020