Recall Z-1324-2025 ยท JAK System, X-Ray, Tomography, Computed

Philips Medical Systems Nederland B.V.

Spectral CT on Rails, Software Version Number 5.1.0, Model number: 728334;

DateMar 12, 2025 (initiated Feb 28, 2025)
StatusOpen, Classified
ReasonMultiple problems identified with the software version leading to various scanning and image issues, and unintended device movement.
Root causeUnder Investigation by firm
ActionOn February 28, 2025 URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. Actions that should be taken by the customer / user in order to prevent risks for patients or users. " You may continue to use your system(s) in accordance with the intended use and by following the recommendation listed below. " When you see the warning message No communication with table and C-arm. Avoid collisions, Proceed with caution., acknowledge the message by selecting OK and continue using the CT while observing with extra care, collision with the Gantry and C-arm can be avoided. " To protect the pat
Quantity1 unit
DistributionDomestic: MN
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K212875Spectral CT on Rails
Philips Medical Systems Nederland B.V., May 13, 2022