Recall Z-1508-2025 ยท DXN System, Measurement, Blood-Pressure, Non-Invasive

PYRAMES INC

Boppli: Bedside Device Kit, REF: 54-0002; Sensor Band, REF: 51-0007

DateApr 2, 2025 (initiated Feb 24, 2025)
StatusOpen, Classified
ReasonSingle-use battery-powered blood pressure monitor, for use on neonates and infants, due to vibratory noise, may not adequately detect changes in blood pressure per specification. Containment actions: 1) Ventilator lines not to touch the isolette (e.g. mattress, side rails, pillow) or patient, 2) Move affected monitor to patient's foot.
Root causeSoftware design
ActionThe firm sent a MEDICAL DEVICE CORRECTIONS NOTTICE" dated 3/5/2025 to its customers describing the product, problem and actions to be taken. Starting on 2/24/2025, the firm started providing training to users to inform them of the following: Signal may be green, but high frequency noise appears superimposed on pulse waveform signal. Containment actions: 1) Ensure that ventilator lines do not physically touch the isolette (e.g. mattress, side rails, pillow) or baby. 2) Move Boppli to patient's foot. On 3/7/2025, correction notices were emailed to customers who in addition to the above we
Quantity35
DistributionUS distribution to states of: MA and CA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K223873Boppli Infant Blood Pressure Monitor
Pyrames, Inc., Sep 29, 2023