Recall Z-1370-2025 ยท OJH Orthopedic Tray
MEDLINE INDUSTRIES, LP - Northfield
Proxima Drape labeled as ORTHO TRAUMA CDS, TOTAL KNEE AUX PACK, TOTAL HIP PACK, TOTAL HIP, TOTAL KNEE, SPINE. Sterile surgical drape.
| Date | Mar 14, 2025 (initiated Feb 13, 2025) |
| Status | Open, Classified |
| Reason | Potential breach in pouch packaging which could lead to loss of sterility. |
| Root cause | Packaging process control |
| Action | Medline distributed an IMMEDIATE ACTION REQUIRED MEDICAL DEVICE RECALL notification, dated 2/13/25, to consignees who purchased individually packaged Proxima sterile surgical drapes and kits that contained Proxima sterile surgical drapes. Consignees that purchased kits containing recalled product are to immediately check stock for affected kits and place them in quarantine. Consignees are to complete the provided response form to receive over-labels to be applied to quarantined kits with additional instructions to remove the affected components prior to use of the kit. Consignees that purcha |
| Quantity | 1,118 units |
| Distribution | US Nationwide distribution. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.