Recall Z-1403-2025 ยท MBF Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer, Uncemented

ZIMMER ORTHOPEDICS MANFACTURING LIMITED-GALWAY

Comprehensive Shoulder Stem, Comprehensive Micro Stem, Comprehensive Mini Stem; uncemented shoulder prosthesis, Model Nos. 113605, 113606, 113607, 113627

DateMar 20, 2025 (initiated Mar 3, 2025)
StatusOpen, Classified
ReasonFive complaints received where surgeon was unable to remove and/or connect the inserter with the stem during surgery due to excess material remaining within the taper. Issue may lead to clinically significant extension of surgery to find a replacement.
Root causeNonconforming Material/Component
ActionOn March 3, 2025, the firm notified affected customers via Urgent Medical Device Recall letters. Customers were informed of the excess metal on the inner taper of the stem due to a chipped manufacturing tool. Customers were asked to notify all affected persons and assist their Zimmer Biomet sales representative in quarantining all affected product on hand. The products will be removed from your facility.
Quantity90 US; 136 OUS
DistributionUS distribution nationwide. International distribution to Dominican Republic, Canada, India, Australia, Japan, EMEA, Korea, Taiwan, and Hong Kong.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K193038Comprehensive Shoulder System
Biomet Manufacturing Corp, Oct 28, 2020