ClearanceBrief
Recall Z-1480-2025 ยท NHB Polymer, Ear, Nose And Throat, Synthetic, Absorbable

Stryker Corporation

Nasopore 4cm Standard 2PK fragmentable nasal dressing, Catalog Number 5400-212-004S and Nasopore 4cm Firm 2PK Catalog Number 5400-222-004S

DateMar 28, 2025 (initiated Feb 25, 2025)
StatusOpen, Classified
ReasonThere is a potential for blister seals on the product to present a bubble on the seal area of the packaging. This seal acts as a sterile barrier and, in this event, indicates that the sterility seal has been breached.
Root causeUnder Investigation by firm
ActionAn Urgent Medical Device Recall notification letter dated 2/24/25 was sent to customers. Actions to be taken 1. Immediately review your inventory to locate and quarantine any affected products at your facility. 2. Return the enclosed Business Reply Form (BRF), even if the affected product is no longer in inventory, to the email Instruments.Recalls@stryker.com. 3. Upon receipt of the completed BRF, Stryker will provide a shipping label to return recalled product on-hand. 4. A replacement will be provided upon receipt of the recalled product. 5. Maintain awareness of this communication
Quantity182,344 total units
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K070715NASOPORE EAR
Polyganics B.V.,, May 7, 2007