Checkpoint Surgical Inc
Brand Name: Checkpoint Guardian Intraoperative Lead Product Name: Checkpoint Guardian Intraoperative Lead, Small Model/Catalog Number: REF# 9524 (Small) Product Description: The Checkpoint Guardian Intraoperative Leads are single-patient disposable accessories for providing stimulus to a targeted nerve. The accessories are intended for use with any device from the Checkpoint Stimulator/Locator Fam
| Date | Mar 26, 2025 (initiated Feb 24, 2025) |
| Status | Open, Classified |
| Reason | Adhesive on the back of the lead may not fully encapsulate the metallic portion of the wire and button subassembly, potentially leading to exposed metal. There is a risk of electrical current leakage, which may lead to inconsistent or confusing muscle responses, potentially compromising the effectiveness of the intended use as the surgeon could take an incorrect clinical action. |
| Root cause | Process control |
| Action | The recalling firm alerted affected hospitals by visit on February 24, 2025. Customers were asked to check inventory for affected lot numbers and to return the affected inventory for replacement. Notifications were issued on February 26, 2025. |
| Quantity | 135 |
| Distribution | US Nationwide distribution. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K212355 | Checkpoint Guardian Intraoperative Lead, Medium, Checkpoint Guardian Intraoperative Lead, Large Checkpoint Surgical, Aug 26, 2021 |