Recall Z-1487-2025 ยท LLZ System, Image Processing, Radiological

GE Healthcare

GE Healthcare Centricity Centricity PACS-IW with Universal Viewer, Model Numbers: 1) 2049587-015; 2) 2066908-086; 3) 2066908-136; 4) 2104867-045.

DateMar 28, 2025 (initiated Feb 19, 2025)
StatusOpen, Classified
ReasonThere is a potential security vulnerability in Centricity Universal Viewer (UV), Centricity PACS-IW (PACS-IW), Centricity Radiology RA600 (RA600), & Centricity Cardiology CA1000 (CA1000) where the service login credentials are able to be identified which could allow a malicious actor with these credentials to access the system and potentially manipulate patient data.
Root causeSoftware design
ActionGE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 02/19/2025 by letter delivered using traceable means. The notice explained the issue, safety issue, and requested the following actions be taken: "You can continue to use your device. Please ensure all potential users in your facility are made aware of this safety notification. Please ensure you have implemented the actions below: 1: Ensure your hospital network is secure and 2: Outside of the hospital network implement additional safeguards such as a VPN. Please retain this document for your records. Pleas
Quantity4 units
DistributionWorldwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.